REMOTE OPERATION
Command Syntax
7/01 LDC-3700 Series 49
CHAPTER 3
Note:If the GPIB times out while waiting for a response, either set the GPIB time−out longer
or use SRQ generated interrupts in your program. See your GPIB manual for time−out
configuration or SRQ programming setup. The *OPC, *OPC?, and *WAI commands should
not be used inside a calibration routine.
Output Off Registers
The Output Off Enable Registers allow you to determine which conditions and events in the TEC
and LASER controllers can cause their outputs to be turned off. These registers are configured in a
manner which is similar to the status reporting registers. However, their outputs are not reported in
the Status Byte Register. Rather, they go to the hardware which controls the output switching for
that function (LASER or TEC). The events and conditions which may be set to cause the TEC and
LASER outputs to be turned off are shown in Figure 3.6 on page 49 and Figure 3.7 on page 50.
The default (factory) settings for these registers are shown in Table 3.4. These settings are not
effected by the *PSC (Power−On Status Clear) command.
Figure 3.6 LASER Output Off Register
Table 3.4 Default Settings for Output Off Registers
LASER Output Off Register TEC Output Off Register
0− disabled 8− N/A 0− disabled 8− enabled
1− disabled 9− disabled 1− disabled 9− disabled
2− N/A 10− disabled 2− N/A 10− enabled
3− enabled 11− enabled 3− enabled 11− N/A
4− N/A 12− N/A 4− N/A 12− N/A
5− N/A 13− N/A 5− enabled 13− N/A
6− N/A 14− N/A 6− enabled 14− N/A
7− N/A 15− N/A 7− enabled 15− N/A
76543210
76
54
3
2
10
15
14
13 12
11 10
8
9
101112131415
9
8
Logical
OR
Turn Output Off
LASER Output Off
Register
LASER Output Off
Enable Register
LASer:ENABle:OUTOFF <nrf>
LASer:ENABle:OUTOFF?
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&