7.8.2 Termination of Charging Session Due to IPG
Temperature
To ensure your safety while charging your OPTIMIZER
Integra CCM-D IPG, the temperature of your IPG is
monitored during the charging process. If the reported
temperature of your OPTIMIZER Integra CCM-D IPG at
the beginning of the charging session is outside the
acceptable temperature range or if the temperature of
your implanted OPTIMIZER Integra CCM-D IPG
remaining consistently high for more than 10 minutes
while it is being charged, then your Guardio Charger will
emit 3 long beeping tones and display the Charging IPG
Temperature Error screen, indicated by a thermometer
icon in the center of the screen (see Figure 21). Your
Charger will then shut off. If this should occur, please wait
for approximately 10 minutes before initiating a new
charging session.
Figure 21: Charging IPG Temperature Error Screen
7.8.3 Termination of Charging Session Due to
Charging IPG Timeout
If the duration of the charging session exceeds 5 hours,
your Guardio Charger will emit 3 long beeping tones and
display the Charging IPG Timeout Error screen, indicated
by a flashing hourglass icon in the center of the screen
(see Figure 22). Your Charger will then shut off. If this
should occur, please wait for approximately 20 minutes
before initiating a new charging session.
Figure 22: Charging IPG Timeout Error Screen