7.8 Termination of the Charging Session
7.8.1 Early Termination of Charging Session
To terminate a charging session before it has been
completed, press and hold the Power Button down for
one second and then release it. Your Guardio Charger will
emit 3 short beeping tones and display the Charge
Session Cancelation screen, indicated by a flashing
universal power icon in the center of the screen. See
Figure 20.
Figure 20: Charge Session Cancelation Screen
Alternatively, you can remove the charging wand from
your implant site, which will cause your Guardio Charger
to time out and shut off automatically.
Note: If you wish to resume charging your OPTIMIZER
Integra CCM-D IPG after you terminate a charging
session, please wait for approximately 10 minutes before
initiating a new charging session to allow the temperature
of your OPTIMIZER Integra CCM-D IPG to return to its
baseline temperature.