Quality of Service (QoS) for Communications between
the Guardio Charger and the OPTIMIZER Integra
CCM-D IPG
MedRadio in the MICS sub-band (402 to 405 MHz)
wireless technology enables communication between the
OPTIMIZER Integra CCM-D IPG and the Guardio
Charger. The requirements for the Quality of Service
(QoS) vary depending on the use environment (operating
room, recovery room, clinic, and home environment).
The Guardio Charger will begin by displaying the IPG
Data Download and IPG Data Download Success
screens:
After the data download has been completed, the
Charging IPG Status screen is displayed by the Guardio
Charger:
The Coupling Level icon ( ), whose number of
illuminated bars is proportional to the proximity of the
charging wand to the implanted OPTIMIZER Integra
CCM-D IPG, is indicative of the Quality of Service (QoS)
for the transcutaneous energy transmission wireless link.
The charging wand should be repositioned until at least 2
bars of the Coupling Level icon are illuminated, indicating
sufficient QoS for charging the OPTIMIZER Integra CCM-
D IPG.