57
Guardio Charger Wireless Nominal Specifications
Characteristic
Nominal
MICS MedRadio
Frequency Band
402 405 MHz Medical Implant
Communication Service (MICS)
Medical Device Radio
Communication Service
(MedRadio)
Bandwidth
240 kHz
Modulation
FSK
Radiated Power
-20.6 dBm EIRP
Range
0 to at least 1.5 m
Transcutaneous Energy Transfer
Frequency Band
13.56 MHz
Industrial, Scientific, and Medical
radio band (ISM)
Modulation
Amplitude (slow to optimize
coupling)
Radiated Power
< 0.6 W reactive near-field
Range
5 mm to 40 mm
Recharge Channel Communication
Frequency Band
13.56 MHz ± 9.2 ppm
Industrial, Scientific, and Medical
radio band (ISM)
Bandwidth
< 0.014 MHz
Modulation
PPM
Radiated Power
-6.93 dBm EIRP
Range
5 mm to 40 mm

Table of Contents