33
The battery inside your Guardio Charger is expected to have a
service life of at least 5 years. If your Guardio Charger is unable to
fully charge your OPTIMIZER Integra CCM-D IPG after the
Charger’s internal battery has been fully charged, please call the
24-hour Support Hotline (866-312-5370) to obtain a replacement
Guardio Charger.
7.15 Storage and Handling
Your Guardio Charger System should not be exposed to
excessively hot or cold conditions. Store your Guardio Charger
System in a cool, dry place, with your Guardio Charger connected
to its AC Adapter and the AC Adapter plugged into the wall outlet.
Do not leave your Guardio Charger System in your car or
outdoors for extended periods of time. The sensitive electronics of
your Guardio Charger System can be damaged by temperature
extremes, particularly high heat.
For proper operation, your Guardio Charger should be used only
under the following environmental conditions:
• Ambient Temperature: 50°F to 81°F
• Relative Humidity: 20% to 75%
• Atmospheric Pressure: 20.73 inHg to 31.39 inHg
Note: Your Guardio Charger is designed for indoor use.
If necessary, move to a location that meets these conditions prior
to using your Guardio Charger.
7.16 Disposal
If your Guardio Charger is no longer needed, you may return it to
your doctor’s office.
Warning: DO NOT discard your Guardio Charger in the trash bin.
Your Guardio Charger contains Lithium-ion batteries as
well as non-RoHS components. If disposal of your
Guardio Charger is necessary, properly dispose of it in
accordance with local regulations governing the
disposal of such material.