39
ISO 14117:2019
Clause 4.8 -
Protection from
AC magnetic field
exposure in the
range of 1 kHz to
140 kHz
Per standard
Does not exhibit
malfunction which
persists after the
removal from the
field per standard
Seek the advice of your
physician or other qualified
health provider regarding
Environmental
Conditions, Industrial
Machinery, and Home
Appliances.
Exercise caution in
the vicinity of
equipment that
generates strong AC
magnetic fields.
Do not enter an area
with posted warnings
advising pacemaker
patients (or patients
with other types of
implantable devices)
not to approach.
ISO 14117:2019
Clause 4.9 - Test
requirements for
the frequency
range of 385 MHz
≤ ƒ ≤ 3000 MHz
Per standard
Functions as it did
before the test
without further
adjustment after
application of the
test signal per
standard
Seek the advice of your
physician or other qualified
health provider regarding
Transmitting Devices and
Cellular and Mobile
Phones
Exercise caution in
the vicinity of
equipment that
generates strong
radio-frequency fields.
Do not enter an area
with posted warnings
advising pacemaker
patients (or patients
with other types of
implantable devices)
not to approach.
Interference may
occur in the vicinity of
equipment marked
with the following
symbol: