EasyManuals Logo

IMT Analytics PF-300 User Manual

Default Icon
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #12 background imageLoading...
Page #12 background image
Specifications
12
4
IMT Analytics AG
Gas standard Temperature Pressure Relative
humidity
API Standard Conditions 15/1013 15°C (60°F) 1013.25 mbar
(14.7 psia)
0%
Cummings Standard 25/991 25°C (77°F) 991 mbar
(500 ft altitude)
0%
20°C/1013 mbar 20/1013 20°C (68°F) 1013.25 mbar
(760 mmHg)
0%
Normal Temperature and
Pressure
NTPD 20.0°C (68°F) 1013.25 mbar
(760mmHg)
0%
Normal Temperature and
Pressure, Saturated
NTPS 20.0°C (68°F) 1013.25 mbar
(760mmHg)
100%
In this User manual the unit sL/min is based on ambient conditions of
0°C and 1013 mbar (DIN 1343).
Please refer to Appendix B: Measurement parameters and units.
There you will also find the conversion factors for the units of meas-
urement
Input voltage of the power supply unit: 100 240 VAC, 50 60 Hz
Supply voltage: 15 V DC
Power input: 25 V A
Operating time in battery operation: 3 hours
Operating time in battery operation
with MultiGasAnalyser 2 hours
Charging the battery
A complete charging process takes 8 h. The service life of the battery is extended if
the battery is charged completely only after a prompt by the device.
Thedevice indicates visually and audiblywhen the battery has to be
charged. Please do not store the battery in the depleted state. Cau-
tion:depletion can damage the battery beyond repair!
4.3 Power supply
4.4 Battery operation

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the IMT Analytics PF-300 and is the answer not in the manual?

IMT Analytics PF-300 Specifications

General IconGeneral
BrandIMT Analytics
ModelPF-300
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals