Specifications
12
4
IMT Analytics AG
Gas standard Temperature Pressure Relative
humidity
API Standard Conditions 15/1013 15°C (60°F) 1013.25 mbar
(14.7 psia)
0%
Cummings Standard 25/991 25°C (77°F) 991 mbar
(500 ft altitude)
0%
20°C/1013 mbar 20/1013 20°C (68°F) 1013.25 mbar
(760 mmHg)
0%
Normal Temperature and
Pressure
NTPD 20.0°C (68°F) 1013.25 mbar
(760mmHg)
0%
Normal Temperature and
Pressure, Saturated
NTPS 20.0°C (68°F) 1013.25 mbar
(760mmHg)
100%
In this User manual the unit sL/min is based on ambient conditions of
0°C and 1013 mbar (DIN 1343).
Please refer to Appendix B: Measurement parameters and units.
There you will also find the conversion factors for the units of meas-
urement
Input voltage of the power supply unit: 100 240 VAC, 50 60 Hz
Supply voltage: 15 V DC
Power input: 25 V A
Operating time in battery operation: 3 hours
Operating time in battery operation
with MultiGasAnalyser 2 hours
Charging the battery
A complete charging process takes 8 h. The service life of the battery is extended if
the battery is charged completely only after a prompt by the device.
Thedevice indicates visually and audiblywhen the battery has to be
charged. Please do not store the battery in the depleted state. Cau-
tion:depletion can damage the battery beyond repair!
4.3 Power supply
4.4 Battery operation