MultiGasAnalyser OR-703
47
7
IMT Analytics AG
MultiGasAnalyser is non-sterile. Autoclaving, sterilisation and immersion in liquid can
cause serious damage to the sensor. The sensor can be cleaned with a cloth mois-
tened with ethanol or isopropyl alcohol.
The airway adapter must be changed at least every 12 months. If MultiGasAnalyser
is used in a sterile system, a new sterile adapter must be fitted.
Gas measurements should be checked regularly using a reference measuring instru-
ment.
For this purpose IMT Analytics AG offers a service to recertify the measuring accuracy
of the sensor.
Physical data Dimensions (L × W × H) 38 × 37 × 34 mm
1.49 × 1.45 × 134 inches
Weight <25 g (not including cable)
Cable length 2.50 m ±0.02
Ambient conditions Operating temperature 10–40°C, 50–104°F
Storage temperature -20–50°C, -4–122°F
Air humidity (operation) 10–95% RH, non-condensing
Air humidity (storage) 5 –100% RH, condensing
Atm. pressure (operation) 700 –1200 hPa
Accuracy
specifications
(under standard
conditions)
Gas Range Tolerance
CO
2
0–10% ±(0.2% ABS + 2% REL)
10–20% ±(0.3% ABS + 4% REL)
N
2
O 0–100% ±(2% ABS + 2% REL)
HAL, ISO, ENF 0–8% ±(0.15% ABS + 5% REL)
10–15% ±(0.2% ABS + 10% REL)
DES 0–22% ±(0.15% ABS + 5% REL)
Rise times
(@ 10 L/min)
CO
2
< 90 ms
N
2
O, HAL, ISO, ENF, SEV, DES < 300 ms
Monitoring Numerical measurement data and real time graph representa-
tion with
FlowLab software.
Deviations from gas setting stated. For example, 50 vol% helium typically reduces
the CO
2
values by 6%. This means that a measured mixture consisting of 5.0 vol%
CO
2
and 50 vol% helium is equivalent to a measured concentration of (1-0.06) * 5.0
vol% = 4.7 vol% CO
2
.
7.8 Servicing and care
7.9 Technical specifications