Filter Replacement
Inspect the air filter regularly. Replace the filter when it appears dirty, shows signs
of wear, after 90 hours of use, or after 90 days of use, whichever comes first.
If a nebulizer is used along with the inCourage System, it is recommended
that the filters be replaced monthly. Failure to change filter could impede
device performance.
Fuse Replacement
1. Remove the black fuse holder located on the back of
the device using a small flat head screwdriver.
(Image A & B)
2. Replace both fuses with 5 A, 125 VAC,
5 mm x 20 mm, slow blow, 10000 A interruption
rated fuses. (Image C)
3. Insert replacement fuses into holder and place back
into unit until you hear a click. (Image D)
Filter panel is located on the right side of the device with the
RespirTech logo.
1. Unplug the PTU.
2. Grasp bottom of side panel labeled RespirTech. Gently
pull away from unit until you hear a click. (Image A)
3. Pull panel down until it is completely disconnected from
the unit. (Image B)
4. Pull out existing filter and replace with new filter. The
filter must go behind the four small tabs and in front of
the support grate. (Image C).
5. To reconnect the panel, align the tabs and gently push
up and in at the bottom until you hear a click for a
secure fit.
A
C
A
C
B
D
B
NOTE: If service is required, please call a RespirTech Customer Support
Representative at 1-800-793-1261.