EasyManua.ls Logo

inCourage RespirTech - Specifications; Product Specifications

inCourage RespirTech
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Principles of Operation
The inCourage
®
Airway Clearance Therapy utilizes three main components:
1. Pulsating Therapy Unit (PTU)
2. Inflatable Vest
3. Interconnecting Hoses
These components work in conjunction with each other to deliver pulsating positive
pressure air pulses to the patient known as High Frequency Chest Compression
(HFCC). The rhythmic inflation and deflation of the vest against the patient’s chest aids
in the mobilization and clearance of bronchial secretions.
Product Specifications
System: inCourage Airway Clearance Therapy, Model ICS-1M-US-A
Manufactured By:
Mode of Operation: Continuous Use
Operating Temperature: 10º C to 33º C (50º F to 92º F) Ambient Temperature
Storage/Transport: -25º C to 85º C (-13º F to 185º F)
Operating Humidity: 20% to 80% RH (non-condensing)
PTU Length: 13.5”
PTU Width: 9”
PTU Height: 13”
PTU Weight: 17.5 lbs.
Input Source: 120 VAC RMS, 50/60Hz, Single Phase
Wattage: 500 Watts
Fuses: 5 A, 125 VAC, 5 mm x 20 mm, slow blow , 10000 A interruption rating
Vest Specifications
1050 Denier Nylon outer shell, PVC- coated
200 Denier Nylon inner shell, PVC- coated
Nylon auto-adjusting securement straps with quick-release connectors
Respiratory Technologies, Inc.
2896 Centre Pointe Drive, St. Paul, MN 55113
SPECIFICATIONS
28
inCourage
®
Airway Clearance Therapy | INSTRUCTION MANUAL