6 | Operation CU1000 INFICON
18 / 36 CU1000-operating-instructions-jina54en1-06-(1901)
View settings
Control unit Settings > Favorites > Favorite 1 (2, 3)
Display of messages
on the touch screen
Warnings and error messages can be displayed on the touch screen.
Off
On
Control unit Settings > Set up > Control unit > Messages > Show
warnings
Show calibration note
Suppress or allow the calibration note with the following content:
Leak rate of the applied calibration leak
No calibration should take place during the first 20 mins
OFF (suppressed)
ON (allowed)
Control unit Settings > Set up > Control unit > Messages > Show
calibration notes
Show calibration
request
The calibration request can be allowed or suppressed.
OFF (suppressed)
ON (allowed)
Control unit Settings > Set up > Control unit > Messages > Show
calibration request
Display of messages
on the touch screen
Warnings and error messages can be displayed on the touch screen.
Off
On
Control unit Settings > Set up > Control unit > Messages > Show
warnings
Setting the audio alarm
Volume of the headphones or active speaker
--- No sound
Proportional: The frequency of the audible signal is proportional to the bar graph
display or diagram height. The frequency range is 300 Hz to 3300 Hz.
Setpoint: The pitch is proportional to the leak rate. The signal sounds if the leak rate
exceeds the selected trigger value.