Supply
DANGER
The gauge may only be connected to power sup-
plies, instruments, or control devices that conform
to the requirements of a grounded protective extra-
low voltage (PELV) and limited power source
(LPS), Class 2. The connection to the gauge has to
be fused.
2)
Supply voltage
at the gauge
3)
+14.5+30 V (dc)
Ripple
1 V
pp
Power consumption MxG50x
≤2 W
Fuse to be connected
2)
1 AT
High voltage in the measuring chamber
Ignition voltage
Operating voltage
≤4.5 kV
≤3.3 kV
3Mxx-x0x-xxxx
3Mxx-x1x-xxxx
high current
low current
Electrical connection
3Mxx-xxx-00xx
3Mxx-xxx-0
1
xx
1)
FCC 68, 8-pin
D-sub, 9-pin
Sensor cable
FCC connector
D-sub connector
8-pin, shielded
9-pin, shielded
Cable length (FCC only) ≤50 m
(0.14 mm
2
/conductor)
2)
INFICON controllers fulfill this requirement.
3)
The minimum voltage of the power supply unit must be increased
proportionally to the length of the sensor cable.