Section 2: Safety
6 of 34
2.8 Warnings
This equipment is for use only by qualified medical professionals trained in
the particular technique to be performed.
Only Handpiece manufactured or approved by InMode Ltd. should be used
with InMode System with Morpheus8 Body Handpiece.
Connect the power cord to a properly polarized and grounded power
source with the frequency and voltage characteristics that match those
listed on the back of the unit.
Connect the System power cord to a properly grounded receptacle. Do not
use power plug adapters.
Always turn off and unplug the device before cleaning.
The patient should not come into contact with metal parts which are
earthed or which have an appreciable capacitance to earth. The use of
antistatic sheeting is recommended for this purpose. Treatment bed or
chair should not be electric.
Use the lowest output setting necessary to achieve the desired treatment
effect. The higher RF is applied, the greater the possibility of unintended
thermal damage.
Failure of the equipment could result in an unintended increase of output
power.
The cables of the Handpiece should be positioned in such a way that
contact with the PATIENT or other leads is avoided.
Fire / Explosion Hazard - The following substances will contribute to
increased fire and explosion hazards in the operating room:
Flammable substances (such as alcohol-based skin prepping agents and
tinctures).
Naturally occurring flammable gases which may accumulate in body
cavities such as the bowel.
Oxygen enriched atmospheres.
Oxidizing agents (such as nitrous oxide [N2O] atmospheres).