▪ If alcohol is used for cleaning and disinfecting, it must be allowed to dry
thoroughly before the System is used.
2.6 Safety Features of the System
The System incorporates the following safety features. All personnel operating the
System should be familiar with these features.
▪ The System has unique password to avoid device operation by non-authorized
personnel.
▪ The power electronics cannot be activated unless the applicator and Footswitch
have been connected to the System.
▪ An audible tone indicates energy activation.
▪ During activation, the System performs a self-test of the hardware.
▪ Hardware is tested every 10ms to ensure proper operation of electrical circuit.
▪ The System starts at a low setting.
2.7 Safe use of the Active Accessories
▪ Examine the connection of the Handpiece through the connectors to the System
before using. Ensure that the accessory functions as intended. Improper
connection may result in arcs and sparks, accessory malfunction, or unintended
treatment effects.
▪ Do not wrap the Handpiece cords around metal objects. It may induce current
that could lead to electrical shocks, fire or injury to the patient or personnel.
Do not connect a wet accessory to the System.
Do not immerse the applicator under water at any time.