Section 8: System Specifications
27 of 34
Section 8: System Specifications
Input Power
Main Line Frequency (nominal)
50-60Hz
Input Voltage (nominal)
100-240VAC
Input Current (rms)
2A
Operating Parameters
Ambient Temperature Range
15 30°C [59 86°F]
Relative Humidity
30% to 80%, non-condensing
Atmospheric Pressure
90 - 110kPa
Warm-up Time
If transported or stored at temperatures outside
the operating temperature range, allow one hour
for the device to reach room temperature before
use.
Transport and Storage
Ambient Temperature Range
-20 65
o
C [-4 14°F]
Relative Humidity
0% to 80%, non-condensing
Atmospheric Pressure
50 to 110kPa
Dimensions
System
46cm W x 46cm
D x 100cm H
[18.2’’ W x 18.2’’ D x
40’’ H]
Handpiece Cable
280cm L
[100`` L]
Weight
System
15.000kg
[33.069lb]
Morpheus8 Body Applicator
0.400kg
[0.243lb]
Morpheus8 Body Output Parameters
Maximum Output Power
65[W]
Frequency
1MHz
Crest Factor (Rated Load)
1.4± 2%

Related product manuals