Display Icon Description What To Do
Battery HOT
The battery has exceeded tem-
perature limit while concentrator
is running on battery power. The
concentrator will shut down and
stop producing oxygen.
If possible, move concentrator to a
cooler location, then turn power off
and back on. Ensure air intake and
outlet vents have clear access and
particle filters are clean. If condition
persists, switch to external power
or a backup source of oxygen and
contact your equipment provider.
System HOT
Concentrator temperature is too
high. The concentrator will shut
down and stop producing oxygen.
Ensure air intake and outlet vents
have clear access and particle
filters are clean. If condition per-
sists, switch to a backup source of
oxygen and contact your equipment
provider.
Sensor fail
The concentrator’s oxygen sensor
has malfunctioned.
You may continue to use the con-
centrator. If the condition persists,
contact your equipment provider.
System COLD
The system is cold (<2˚C). The con-
centrator will shut down and stop
producing oxygen.
Move to a warmer environment to
allow the unit to warm up before
starting it. If condition persists,
switch to a backup source of oxy-
gen and contact your equipment
provider.
System Error
The concentrator has stopped pro-
ducing oxygen and is shutting down.
Switch to backup oxygen source
and contact your equipment
provider.
9. TROUBLESHOOTING
Problem Possible Cause Recommended Solution
Any problem accompanied by infor-
mation on concentrator display, indi-
cator lights and/or audible signals
Refer to device icon & alarm
glossary
Refer to device icon & alarm
glossary

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Inogen Rove 6 and is the answer not in the manual?

Inogen Rove 6 Specifications

General IconGeneral
Oxygen Concentration90% (+6% / -3%)
Flow Settings1-6
Power OptionsAC, DC, and battery
Battery DurationUp to 6.5 hours
Noise Level38 dBA

Related product manuals