18. Alarms, User Alerts and User Actions continued
Alarm On-Screen
Message(s)
Alarm
Description
Actions to Clear/
Silence
Cause(s) Actions
Full screen warning.
Visual : AMBER (LOW)
Audible : LOW
Internal device
temperature
measures >85°C
When this alarm state
is initiated, the unit will
automatically shut-down
in 1 minute.
Operator can press the
power button to turn
the unit OFF before the
unit automatically shuts
down.
Operator must allow the
device to cool and ensure
air vents are not blocked
and then restart the
device.
Failure of internal
components.
Failure to adequately
cool (e.g. blocked
vents).
Cut all power to
warming system.
Alert sounds
for 60 seconds
followed by system
shutdown.
Contact technical
support for repair.
Full screen warning.
Visual : AMBER (LOW)
Audible : LOW
Internal fan failure. When this alarm state
is initiated, the unit will
automatically shut-down
in 1 minute.
Operator can press the
power button to turn
the unit OFF before
the unit automatically
shuts down.
Operator may attempt to
power cycle the device.
Failure of internal fan.
Failure of internal
components.
Cut all power to
warming system.
Alert sounds
for 60 seconds
followed by system
shutdown.
Contact technical
support for repair.
Full screen warning.
Visual : AMBER (LOW)
Audible : LOW
Battery Critical. When this alarm state
is initiated, the unit will
automatically shut-down
in 1 minute.
Operator can press the
power button to turn
the unit OFF before
the unit automatically
shuts down.
Operator must connect
external power and then
restart the device.
Battery is critically
depleted and no
external power supply
is connected.
Failure of battery
status monitoring
circuit.
Loss of supply mains
when battery is not
yet appropriately
charged.
Cut all power to
warming system.
Alert sounds
for 60 seconds
followed by system
shutdown.
Contact technical
support for repair.