18. Alarms, User Alerts and User Actions continued
Alarm On-Screen
Message(s)
Alarm
Description
Actions to Clear/
Silence
Cause(s) Actions
Channel over
temperature
flashes amber
with temperature
indication.
Visual : AMBER
(LOW)
Audible : LOW
Temperature
sensor measures
>1°C
over set point
(>5 minutes)
The temperature of the
accessory must drop to
below set point before
heating is restarted on this
channel. If the temperature
of the accessory drops to
<1°C above set point this
alarm will be automatically
cleared.
The alarm active icon will
appear on-screen and
may be used to pause the
audible alarm.
Repeated occurrence of
this alarm may indicate a
fault with this channel or
the attached accessory.
Failure of
warming system.
Cut power to
accessory over
temperature.
AlphaCore
5
Controller
continues
to operate
remaining
accessories.
Contact
technical
support.
Channel disconnected
flashes amber with
disconnect icon.
Visual : AMBER
(LOW)
Audible : LOW
Accessory
disconnected.
User must acknowledge
the alarm by pressing the
disconnect icon on the
active channel, this will
clear the alert.
The alarm active icon will
appear on-screen and
may be used to pause the
audible.
Accessory disconnect
by user (accidentally
or deliberately).
Failure of mattress
circuit.
Failure of AlphaCore
5
Controller .
Cable or Connector
damaged.
Cut power
to accessory
port that has
disconnected.