Patient Warming System Instructions for Use 39
Inspiration Healthcare Limited © 2021 Inspiration Healthcare Limited
W inspiration-healthcare.com
700-307 Revision: 10
18. Alarms, User Alerts and User Actions continued
Alarm On-Screen
Message(s)
Alarm
Description
Actions to Clear/
Silence
Cause(s) Actions
All channels flash
amber with
disconnect icon.
Visual : AMBER
(LOW)
Audible : LOW
Connected
accessories
exceed power
limit.
User is prompted to detach
accessories until the maximum
power output requirement is below
the limit for the currently connected
power source.
When disconnecting accessories
from a unit in an over power limit
state the ALARM-3-DISCONNECT
alert will not sound and any removed
accessories will be automatically
disconnected.
Once the user has disconnected
sufficient accessories to bring the
power requirement within limits the
alert will automatically clear, the
audible alert will be cancelled and
the visual status bar will return to
GREEN indicating normal operation.
Any remaining accessories will return
to normal operation with heating
engaged as required.
The alarm active icon will appear
on-screen and may be used to pause
the audible alarm.
Multiple
accessories are
attached requiring
total output
power greater
than maximum
that is available
from current
power source.
Cut power to
all accessories
until the user
takes action.
Channel with error
flashes amber with
warning triangle icon
Visual : AMBER
(LOW)
Audible : LOW
Accessory
rails
(rail break
detection)
is outside
permitted
range.
User must detach the accessory from
the channel reporting the error.
Contact technical support for repair/
replacement of accessory.
The alarm active icon will appear
on-screen and may be used to pause
the audible.
Failure/
degradation of
accessory internal
construction.
N.B. Typically
this would occur
when a large
mattress has
been repeatedly
folded, damaging
the power rails.
Cut power to
accessory with
out of range
measurement.
AlphaCore
5
Controller
continues
to operate
remaining
accessories.