18. Alarms, User Alerts and User Actions continued
Alarm On-Screen
Message(s)
Alarm
Description
Actions to Clear/
Silence
Cause(s) Actions
Channel connected
flashes amber.
Visual : AMBER
(LOW)
Audible : LOW
New accessory
attached/
detected.
User must acknowledge
the connection alarm
by pressing the flashing
indicator for the newly
connected channel at
which point the alarm is
cleared/silenced.
Warming may then be
applied to the newly
attached accessory
if system is within
operating parameters.
The alarm active icon
will appear on-screen
and may be used to
pause the audible
alarm.
N.B. Power limit
may be breached
by connection of
additional accessories
– see ALERT-2-OVER-
POWER-LIMIT.
System will not
apply warming
to the connected
accessory until
acknowledged
by user.
AlphaCore
5
Controller
continues to
operate remaining
accessories.
No alarm message,
however affected
accessory channel
should be highlighted
in AMBER to indicate
temperature status.
Visual : GREEN
Audible : NONE
Temperature
Sensor measured
>1°C above set
point
(<5 minute).
No user action required
until alert upgraded to
ALARM-3-TEMP-HIGH
after >5 minute over
temperature.
Overshoot on warming
control algorithm.
Failure of warming
system.
Cut power to
accessory over
temperature.
AlphaCore
5
Controller
continues to
operate remaining
accessories.