Patient Warming System Instructions for Use 41
Inspiration Healthcare Limited © 2021 Inspiration Healthcare Limited
W inspiration-healthcare.com
700-307 Revision: 10
18. Alarms, User Alerts and User Actions continued
Alarm On-Screen
Message(s)
Alarm
Description
Actions to Clear/
Silence
Cause(s) Actions
No alarm message,
however affected
accessory should
be highlighted in
BLUE to indicate low
temperature status.
Visual : GREEN
Audible : NONE
Temperature
Sensor
measured
>1°C under
set point.
User must acknowledge
alert to clear audible
indicators. Audible alert
will be paused for 15
minutes.
Accessory will continue
to be highlighted BLUE to
indicate low temperature
status.
Ambient conditions may
limit maximum achievable
temperature.
Failure of AlphaCore
5
Controller to supply
appropriate power.
Failure of mattress.
Failure of temperature
measurement circuit.
Failure of mattress thermistor.
Failure of warming system.
Failure of system to maintain
temperature.
Failure of temperature
measurement circuit.
System
continues to
operate all
connected
accessories.
Low battery icon
below should be
flashing on screen.
Visual : GREEN
Audible : LOW
Battery Low Visual and audible alert
indicators remain active
until device has been
connected to supply mains
or superseded by ALARM-
3-BATTERY-CRITICAL.
After connection to supply
mains this alert will
automatically be cleared.
The audible alert for this
condition may not be
paused.
Battery is depleted (normal
operation).
Failure of battery status
monitoring circuit.
Loss of supply mains
when battery is not yet
appropriately charged.
System
continues to
operate all
connected
accessories
until battery
charge level no
longer supports
warming
operation.
External power and
battery charging
icons disappear.
Visual : GREEN
Audible : Single 2
tone pulse
Loss/
disconnection
of external
power.
No user action required
Single two tone audible
alert sounded to indicate
disconnection of mains.
Mains power supply is
disconnected (either
deliberately or accidentally).
N.B. The device must be
fitted with a suitably charged
internal battery. With no
battery fitted and no external
power source the unit will
simply shut down.
User may
reattach power
supply or
continue to
operate unit
on battery
power subject
to maximum
accessory
load limits.