DOSE IT – Operating instructions V08 Maintenance
24 INTEGRA Biosciences AG
If the system is leaking, you have to check whether the tubing or the pumphead is worn:
Repeat the leak test with a new tubing (only use original INTEGRA tubings, see “7.1
Consumables” on page 27).
• If system is tight, the tubing was worn and the problem is solved.
• If the system is still leaking, the pumphead is worn (e.g. lever loose, mechanical play)
and need to be replaced, see “7.2 Spare part” on page 28
.
N
OTE
Work around until a new pumphead is available: Put a rubber band (5 to
8 mm wide) around the pumphead to help holding it fully closed during oper-
ation. Repeat the leak test to confirm that the work around works.