Service Manual
AT235
32
Control whether the tone 226 Hz is audible in the probe tip.
Place probe tip in a 2 cc cavity and check that the colored bar is changing color from blue to green.
The volume shows 2 ml (1.90 to 2.10 ml).
Fig 51