EasyManua.ls Logo

Interacoustics Sera - HearSIM software bundle; Licenses; Adding a new license to Sera; Updating the firmware

Interacoustics Sera
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
D-0114042-D – 2020/03
Sera™ - Additional Information Page 23
Test times using the same detection algorithm and CE-Chirp
®
stimulus are typically less than 30 seconds
456
.
Electrode montage
All Sera
TM
protocols are designed to be use with a nape montage. The electrode montage indicates where
electrodes should be placed for testing and the baby picture displayed on the test screen and on the
preamplifier guides the user where to place the electrodes. The predefined nape electrode montage tells the
system which is the active, reference and ground electrode. Use of an alternative montage (such as a
mastoid montage) may increase test times or produce more refer results.
Result
All protocols will indicate a Pass/Refer result for the test or, if the measurement is stopped before completion
of the test, Incomplete will appear.
4
Almeida, M.G., Sena-Yoshinaga, T.A., Côrtes-Andrade, I.F., Sousa, M.N.C., & Lewis, D.R. (2014). Automated auditory
brainstem responses with CE-Chirp® at different intensity levels. Audiol Commun Res, 19(2), 117-123
.
5
Sena-Yoshinaga, T.A., Almeida, M G., Côrtes-Andrade, I. F., & Lewis, D. R. (2014). Neonatal hearing screening with
automated auditory brainstem response: using different technologies. Audiol Commun Res, 19(1), 19-24.
6
Sena, T.A., Ramos, N., Rodrigues, G.R.I., & Lewis, D.R. (2013). Testing time comparison between two procedures with
new technologies of Automated Auditory Brainstem Response. CoDAS, 25(1), 34-38.
Symbol
Screening result
Pass
Refer
Incomplete

Table of Contents

Other manuals for Interacoustics Sera

Related product manuals