Safety
GuidanceandManufacturer’sDeclaration-ElectromagneticImmunity
Voltagedips,shortinterruptions
andvoltagevariationsonpower
supplyinputlinesIEC61000-4-11
<5%UT
(>95%dipinUT)
for0,5cycle
40%UT
(60%dipinUT)
for5cycles
70%UT
(30%dipinUT)
for25cycles
<5%UT
(>95%dipinUT)
for5sec
<5%UT
(>95%dipinUT)
for0,5cycle
40%UT
(60%dipinUT)
for5cycles
70%UT
(30%dipinUT)
for25cycles
<5%UT
(>95%dipinUT)
for5sec
Mainspowerqualityshouldbe
thatofatypicalcommercialor
hospitalenvironment.Iftheuser
oftheCSBedrequirescontinued
operationduringpowermains
interruptions,itisrecommended
thattheCSBedbepoweredfrom
anuninterruptiblepowersupply
orabattery.
Powerfrequency(50/60Hz)
magneticfieldIEC61000-4-8
3A/m3A/mPowerfrequencymagnetic
fieldsshouldbeatlevels
characteristicofatypicallocation
inatypicalcommercialorhospital
environment.
NOTEUTisthea.c.mainsvoltagepriortoapplicationofthetestlevel.
GuidanceandManufacturer’sDeclaration-ElectromagneticImmunity
TheCSBedisintendedforuseintheelectromagneticenvironmentspecifiedbelow.Thecustomerortheuser
oftheCSBedshouldassurethatitisusedinsuchanenvironment.
ImmunityTest
IEC60601TestLevelComplianceLevelElectromagneticEnvironment–Guidance