Section 2: Safety
5 of 96
Never look directly into the laser aperture at the distal end of the Handpiece, even if
you are wearing safety glasses. Serious eye injury or blindness could result.
Avoid directing the laser beam anywhere other than the calibration port or intended
treatment area. Stray laser light and reflection is always a potential hazard and may
cause serious injury.
Do not treat eyebrows, eyelashes, eyelids or other areas within the bony area
surrounding the orbit. The light emitted by the laser or the IPL is capable of causing
serious eye damage or blindness. For maximum safety, metal eye goggles must be
worn by the patient for all facial treatments when laser is used.
 Ocular Safety Considerations
Identify the laser room clearly by posting approved safety signs in prominent
locations.
Cover all windows to prevent laser light from escaping the laser room.
Restrict entry to the treatment room when the laser or IPL is in use. Allow access
to those personnel both essential to the procedure and well trained in safety
issues.
Never direct the laser beam at anything other than the calibration port or the
intended treatment site.
Never look directly into the laser aperture at the distal end of the Handpiece.
All persons in the treatment room must wear approved laser safety eyewear.
This includes the operator, patient, nurses, and any other persons in the
treatment room.
Do not attempt to remove the plastic shell protective covers on the Handpiece
which could allow exposure to high intensity laser light.
 Electrical and Mechanical Safety
Keep all covers and panels of the System closed. Removing the covers creates a
safety hazard.
Keep hands away from the Applicator during the System start-up.
Perform maintenance procedures when the System is shut down and
disconnected from the power.
The System is grounded through the grounding conductor in the power cable.
This protective grounding is essential for safe operation.