InMode Without Vasculaze Operator Manual
Section 11: BodyFX/MiniFX Treatment
Information
71 of 96
 Treatment Recommendations
Ensure that skin is clean and dry.
No gel allowed!
Follow Device Start-Up Procedure from Section 5.
Set the treatment parameters. Suggested parameters are shown in the table
below:
Handpiece
Treatment
Approach
RF
Power
[W]
Pulse
Width
[Sec]
Cut-Off Skin
Temperature [ºC]
BodyFX
Sensitive
skin
30-40
2
As high as
tolerable
C
o
Starting at 40
BodyFX
Normal skin
35-45
3
As high as
tolerable
C
o
Starting at 40
BodyFX
Resistant
skin
40-50
4
As high as
tolerable
C
o
Starting at 40
MiniFX
Sensitive
area
10
1-3
As high as
tolerable
C
o
Starting at 40
MiniFX
Normal area
15-20
1-3
As high as
tolerable
C
o
Starting at 40
MiniFX
Resistant
area
20-25
1-3
As high as
tolerable
C
o
Starting at 40
Any combination of treatment parameters should be according to skin response and
patient tolerance.
Always start with low settings and observe the skin's reaction and patient
comfort before increasing the RF power, the temperature or pulse width.
Reduce RF Power on thin fat areas and on small areas. When using MiniFX
restrict RF Power to 10-20W, occasionally to 25W.