The patient should not come into contact with metal parts which are
earthed or which have an appreciable capacitance to earth. The use of
antistatic sheeting is recommended for this purpose. Treatment bed or
chair should not be electric!
Use the lowest output setting necessary to achieve the desired treatment
effect. The higher optical or RF energy is applied, the greater the possibility
of unintended thermal damage of skin.
Failure of the equipment could result in an unintended increase of output
power.
The cables of the Handpieces should be positioned in such a way that
contact with the patient or other leads is avoided.
Fire / Explosion Hazard - The following substances will contribute to
increased fire and explosion hazards in the operating room:
Flammable substances (such as alcohol-based skin prepping agents and
tinctures).
Naturally occurring flammable gases which may accumulate in body
cavities such as the bowel.
Oxygen enriched atmospheres.
Oxidizing agents (such as nitrous oxide [N2O] atmospheres).
Endogenous gases.
The optical or RF energy and heating associated with the System can
provide an ignition source. Observe fire precautions at all times. When
using InMode in the same room with any of these substances or gases,
prevent their accumulation or pooling within the area where InMode
procedures are performed.
The operation of the InMode System may adversely influence the operation
of other electronic equipment.
To avoid the risk of electric shock, this equipment must only be connected
to a mains supply with protective earth.