InMode Without Vasculaze Operator Manual
Section 7: DiolazeXL Treatment
Information
46 of 96
Prior to using water-based gel, ensure that the container of ultrasound gel
has not passed the expiration date.
Make a test-spot with Normal/Strong Cooling to see epidermal end-points of
perifollicular erythema/edema. Change to Strong Cooling while treating for
better comfort and safety. Perform the test spot in the area to be treated in a
non-conspicuous site and wait according to the Table below to check the
response.
Test and wait at least 15-30 min and increase or decrease by 2-5 J/cm2 for skin
types I-IV. Wait 2-3 days and increase or decrease by 1-2J/cm2 for skin types V-
VI.
Set treatment parameters. Typical treatment parameters are shown in the
table below; however, an individual adjustment is required:
Skin
Type
]
2
Fluence [J/cm
Pulse
Width
Cooling
755/810
810
810/1064
I
15-30
16-30
18-30
Short
Strong/Normal
II
12-25
14-28
16-30
Short
Strong/Normal
III
10-20
12-25
14-25
Short
Strong/Normal
IV
10-20
12-25
14-25
Long
Strong/Normal
V
6-15
10-20
12-22
Long
Strong
VI
NA
8-16
10-20
Long
Strong
Always start with a low settings level to check patient tolerance. Test, watch
epidermal response, and increase or decrease settings gradually. Darker skin is
safer to treat with Long pulse width. Lighter hair requires higher fluence to
reach the results. Stronger cooling allows better epidermis protection but may
compromise results on very superficial hair e.g. upper lip or Asian skin in
general. Reduce fluence on thin skin and in bone proximity along with a thicker
layer of gel. Consider primarily safety for selection of parameters.
Apply to the treated area ensuring a contact with pressure, depress Footswitch
and press trigger button to deliver DiolazeXL laser pulse.
Move to adjacent site with small overlap and apply next pulses to cover all
treatment area. Pulse Mode may vary from Off (single pulse) or 1, 2 pulses per
second (pps) and Glide (5 pps), restricted by certain fluence, as described in
Device Description in Section 4.
 In a continuous mode choose Pulse Mode of 1, 2 or Glide 5 pps according to
the speed of your hand movement and slide the Handpiece to adjacent site.