TIGER XTL Instrument User Manual V1.0
Pioneering Gas Sensing Technology. Page 54 of 56 ionscience.com
Technical Specifications
Minimum Resolution
0.1 ppm
Detectable Range
0.1 ppm to 5,000 ppm
Response Time
T90 < 2 seconds
Accuracy
± 5% display reading ± one-digit (at calibration point)
Intrinsically Safe Approvals
II 1G Ex ia IIC T4 Ga
Tamb = - 15
o
C ≤Ta ≤+45
o
C (with Lithium ion Battery Pack)
Tamb = - 15
o
C ≤Ta ≤+45
o
C (with Alkaline Battery Pack)
IECEx ITS 22.0025X ITS-I22ATEX35111X
ITS22UKEX0635X
3193491 conforms to UL Std. 913, 61010-1 &
Certified to CAN/CSA Std. C22.2 No. 61010-1
Battery Life Typical operating time of 24 hours
Typical charge time of 8 hours
Alkaline 3 x AA: Typical battery life of 8.5 hours
Lamps
10.6eV krypton (standard)
Data Log
Including date / time: 80,000
Communication
Direct USB 1.1
Calibration
2- and 3-point calibration (via calibration kit accessory)
Alarms
Flashing LED’s amber (low alarm) red (high alarm)
Sounder 95 dBA at 300 mm (12”)
Vibration on alarm
Flow Rate
Temperature
Weight & Dimensions:
≥ 220 ml/min
Operating: -20 to 60
o
C / -4 to 140
o
F (non -intrinsically safe
environment)
Certified to: 15 to +45
o
C / +5 to +113
o
F (Intrinsically Safe
environment)
Storage: -25 to 60
o
C / -5 to 140
o
F
Humidity: 0-99% RH (non-condensing)
Instrument Weight:
870g
Instrument Dimensions:
370 mm (H) x 91 mm (W) x 60 mm (D)

Table of Contents

Related product manuals