Technical Specification
Parameter IT6012B-500-80
Readback stability-8h Voltage
≤0.02% + 0.02%FS
±(of Output+Offset)
Current ≤0.1% + 0.1%FS
Efficiency
92%
Remote Sense Compensa-
tion Voltage
≤5V
Command Response Time
2mS
Power Factor 0.99
Storage Temperature
-10℃~70
Protective Function
OVP, OCP, OPP, UVP, UCP, OTP, Vsense protection
Standard Interface
Standard: USB, CAN, LAN, VCP; optional: GPIB, analog card (in-
cludes RS232), fiber optic socket
Isolation ( output to ground)
1000V
Working Temperature
050
Dimension ( mm) 483mm(W)*801.61mm(D)*151.3mm(H)
Weight( net)
30KG
Note: 1. The instrument needs to be customized under other AC input voltage ranges. For mod-
els equipped with an emergency stop module (including AC contactor), the AC input voltage
range can only be: 342V528V
Load Mode:
Parameter IT6012B-500-80
Rated value ( 0
-50 )
Input voltage
0500V
Input current
080A
Input power
012kW
Input resistance
0.0017500Ω
Min. operating
voltage
2.4V at 80A
Setup Resolution
Voltage
0.01V
Current 0.01A
Power 0.001kW
Resistance 0.001Ω
Read Back
Resolution
Voltage
0.01V
Current 0.01A
Power 0.001kW
Copyright © Itech Electronic Co., Ltd.