IEEE488.2 Commands
Copyright ©ITECH Electronic Co., Ltd. 45
Return parameter:
<NR1>
Reference order:
*CLS *ESE *ESR
*TRG
When the trigger mode of the power supply is BUS order trigger mode, the
order will
Generates a trigger signal. The function is the same as
TRIGger[:IMMediate]
Order syntax:
*TRG
Parameter:
none
Reference order:
TRIG TRIG:SOUR
*SAV
This order can save the parameters of power supply to register. These
parameter contains:
CURR CURR:STEP CURR:TRIG CURR:PROT DISP
OUTP TRIG:SOUR
VOLT VOLT:STEP VOLT:TRIG VOLT:PROT
VOLT:PROT:STAT
Order syntax:
*SAV<NRf>
Parameter:
1~72
Example:
*SAV 3
Reference order:
*RCL