To Adjust the Calibration of an Intelis Gas Meter
Before you perform the Adjust Calibration command, you are required to run the meter on a
prover station. This determines the meter's accuracy, and the results are used during the Adjust
Calibration workflow.
Note: The meter must be out of test mode to perform the Adjust Calibration command. If the
meter is in test mode, perform the Exit Test Mode command before you start this procedure.
Part of this Adjust Calibration workflow includes unsealing the meter. After you complete this
workflow, you must use the separate Seal Meter command to re-seal the meter.
If the meter is operating in mobile mode, you must know the meter's Unlock ID to complete the
Adjust Calibration command.
1. From the Tools menu, select Intelis Gas Meter.
2. Select Test Mode Operations.
3. Select Adjust Calibration.
A warning message appears indicating that Adjusting Calibration will unseal the meter.
4. Select Yes.
5. (Mobile ChoiceConnect mode only) If you are prompted to enter the Unlock ID, enter the
Unlock ID.
6. Select Next. The Parameter screen appears.