EasyManuals Logo
Home>ITRON>Measuring Instruments>Intelis

ITRON Intelis Setup Guide

ITRON Intelis
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #63 background imageLoading...
Page #63 background image
% Proof. 100% open, 100% check
% Accuracy. 100% open, 100% check
% Error. 0% open, 0% check
10. Click Quit or Finish to return to the Tools screen.
11. (Recommended) Rerun the meter on the prover station to verify the meter calibration. It is
expected that you see the results that match the following example:
If the accuracy results are not as expected, repeat the calibration steps.
Note: Prover station accuracy is estimated to be +/- 0.15-0.25%.
To Seal an Intelis Gas Meter
1. From the Tools menu, select Intelis Gas Meter.
2. Select Test Mode Operations.
3. Select Seal Meter.
A progress bar indicates that the meter is sealing.
08 July 2020 815-0154-00 REV 000 Itron, Inc. Page 63 of 66
Intelis Gas Meter Proving and Setup Guide 7 Calibrating the Intelis Gas Meter

Other manuals for ITRON Intelis

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ITRON Intelis and is the answer not in the manual?

ITRON Intelis Specifications

General IconGeneral
BrandITRON
ModelIntelis
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals