9. Product Specifications
9.1 Delivery form
Programat EP 5000
Power cord
Vacuum hose
Calibration test set
Operating Instructions
Programat Firing Tray Kit
Programat USB stick
USB data cable
9.1.2 Recommended accessories
Programat Accessories Set
Temperature Checking Set 2
Vacuum Pump VP4
9.2 Technical Data
Power supply 110–120 V / 50–60Hz
200–240 V / 50–60Hz
Overvoltage category II
Contamination level 2
Tolerated voltage fluctuations +/- 10%
Max. power consumption 12 A at 110–120 V
8.5 A at 200–240 V
Acceptable data for vacuum pump of other manufacturers
Max. output: 250 W / max. leakage current 0.75 mA
Final vacuum: < 50 mbar
Use only tested pumps
Electrical fuses: 110–120 V:
250 V / T 15 A (heating circuit)
250 V / T 5 A (vacuum pump)
200–240 V:
250 V / T 8 A (heating circuit)
250 V / T 3.15 A (vacuum pump)
Dimensions of electrical fuses: 110–120 V:
Diameter 6.3 x 32 mm
200–240 V:
Diameter 5 x 20 mm
Dimensions of the closed furnace:
Depth: 470 mm / width: 305 mm / 400 mm (with cooling tray)
Height: 565 mm
Usable size of the firing chamber: Diameter 80 mm
Height 48 mm
Max. firing temperature : 1200 °C
Weight: Furnace base: 13.5 kg
Furnace head: 7.0 kg
Safety information
The EP 5000 complies with the following guidelines:
EN61010-1:2001 2
nd
Ed.
IEC61010-1:2001 2
nd
Ed.
UL/CSA61010-1:2004 2
nd
Ed.
EN61010-2-010:2003 2
nd
Ed.
IEC61010-2-010:2003 2
nd
Ed.
CSA61010-2-010:2004 2
nd
Ed
Radio protection / electromagnetic compatibility EMC tested
9.3 Acceptable operating conditions
Acceptable ambient temperature range:
+5 °C to +40 °C (+41 °F to +104 °F)
Acceptable humidity range:
80 % maximum relative humidity for temperatures up to 31 °C
(87.8 ° F) gradually decreasing to 50 % relative humidity at 40 °C
(104 °F); condensation excluded.
Acceptable ambient pressure:
The furnace is tested for use at altitudes of up to 2000 m (6562 ft.)
above sea level.
9.4 Acceptable transportation and storage conditions
Acceptable temperature range -20 to +65 °C (-4 °F to +149 °F)
Acceptable humidity range Max. 80 % relative humidity
Acceptable ambient pressure 500 mbar to 1060 mbar
Use only original packaging of the Programat EP 5000 together with
the respective foam material for shipping purposes.