38
The following functions are available:
– Copy programs
– Reset program to factory settings
– Select product logo
– Rename group name
– Rename program name
5.2.8 Pausing the active program
– Press STOP once to pause a running program (green LED flashing). Press the STOP key twice to completely stop the pro-
gram or press START to continue.
– If the program is paused, the display shows a flashing “Pause” and the indicator changes back to the program screen.
– If a program is stopped prematurely, “Vacuum release” is shown in the display during the flooding of the firing
chamber.
5.2.9 Changing the parameters while the program is running
Most program parameters, which have not yet been executed, can be changed while the program is paused. Change the
parameters as described in Chapter 5.2.3.
5.3 Managing programs
Press the button [Admin] in the program view to open
the Admin menu.
5. Operation and Configuration