9. Product Specifications
9.1 Delivery form
Programat P510
Power cord
Vacuum hose
Programat Firing Tray Set 2
Automatic Temperature Checking Set 2 ATK2 (test set)
USB download cable
Operating Instructions
Recommended accessories
Programat Accessories Set
Temperature Checking Set ATK2
Vacuum Pump VP5, VP3 easy
Programat WLAN Kit
9.2 Technical data
Power supply 110 – 120 V / 50 – 60 Hz
200 – 240 V / 50 – 60 Hz
Overvoltage category II
Contamination level 2
Acceptable voltage fluctuations ± 10 %
Max. power consumption 14 A at 100 V
12 A at 110 120 V
8.5 A at 200 – 240 V
Acceptable data for pumps of other manufacturers:
Max. output:
Final vacuum:
250 W / max. leakage current 0.75 mA
< 50 mbar
Only tested pumps should be used.
Electrical fuses
110 – 120 V:
125 V / T15A (heating circuit)
125 V / T5A (vacuum pump)
200 – 240 V:
250 V / T8A (heating circuit)
250 V / T3.15A (vacuum pump)
Dimensions of electrical fuses 110 120 V: Diameter 6.3 x 32 mm
200 240 V: Diameter 5 x 20 mm
Dimensions of the closed furnace Depth: 495 mm
Width: 320 mm / 395 mm (with cooling tray)
Height: 320 mm
Usable size of the firing chamber Diameter: 90 mm
Height: 80 mm
Max. firing temperature 1200 °C
Weight 18.7 kg