8. What If ...
1912 yes Data backup – recover The new software includes program groups which were not yet available at
the time of the data backup. Please check the individual program groups.
The individual program groups are no longer valid.
1920 yes Data backup – recover The data backup could not be performed. Please install a newer software
version.
2754 yes Error self-test An error occurred while checking the infrared camera
2755 yes Error self-test An error occurred while checking the infrared camera
2756 yes Error self-test An error occurred while checking the infrared camera
2757 yes Error self-test An error occurred while checking the infrared camera
2758 yes Error self-test An error occurred while checking the infrared camera
2759 yes Error self-test An error occurred while checking the infrared camera
2760
**
yes Error IRT Check the orientation and positioning of the objects. The objects are
possibly too small or are placed in an unfavourable position on the firing
tray. Check if the protective foil has been removed from the sensor or if
the sensor has been contaminated. Check if a correct firing tray is used.
The height of the firing tray must correspond with the height of the firing
trays recommended for the respective materials by Ivoclar Vivadent AG.
2761
**
yes Error IRT Check the orientation and positioning of the objects. The objects are
possibly too small or are placed in an unfavourable position on the firing
tray. The object was not sufficiently dried in the specified time. Please
repeat the procedure after the object has cooled down.
2762
**
yes Error IRT An error occurred while trying to identify the object. Check if interfering
heat sources are in the visual range of the infrared camera. Sources of
interference are for example: other furnaces, air conditioning systems,
heaters, light bulbs, Bunsen burners, etc. Do not move the objects after the
start of a program.
* Furnace head opens if this error occurs!
** A program in progress is aborted
*** Error cannot be acknowledged; programs cannot be started.
8.2 Additional error messages
Please contact the Ivoclar Vivadent After Sales Service, if one of the following error numbers is being displayed:
25, 29, 43, 44, 45, 46, 47, 48, 56
103, 107, 108, 109, 143, 144, 145, 146, 147
148, 700, 701, 703, 704, 707, 1010, 1011, 1012
1013, 1014, 1015, 1016, 1017, 1018, 1019, 1024, 1025
1026, 1028, 1143, 1144, 1145, 1146, 1147, 1148, 1207
1300, 1301, 1303, 1304, 1305, 1401, 1402, 1500, 1750
1751, 1752, 1753, 2750, 2751, 2752, 2753, 2770