SAFETY
6 Model 3000 Operation Manual
Guidance and manufacturer’s declaration – Electromagnetic immunity
The Model 3000 is intended for use in the electromagnetic environment specified below. The
customer or the user of the Model 3000 should assure that it is used in such an environment.
level
Electromagnetic environment –
guidance
discharge (ESD)
IEC 61000-4-2
±8kV air
±8kV air
Floors should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
Transient/burst
IEC 61000-4-4
supply lines
±1 kV for
supply lines
±1 kV for
Mains power quality should be
that of a typical commercial or
hospital environment.
IEC 61000-4-5
mode
±2 kV common
mode
±2 kV common
Mains power quality should be
that of a typical commercial or
hospital environment.
interruptions, and
voltage variations
on power supply
input lines
IEC61000-4-11
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
) for
5 cycles
70 % U
T
(30 % dip in U
T
) for
25 cycles
<5 % U
T
(>95 % dip in U
T
)
T
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Model 3000 requires
continued operation during power
mains interruptions, it is
recommended that the Model
3000 be powered from an
uninterruptible power supply.
(50/60 Hz)
magnetic field
IEC 61000-4-8