ECG MONITORING
Model 3000 Operation Manual
23
Alarm Limits
1. Press the [SETUP] key from the main menu. The following menu appears.
2. Press the programmable key [LIMITS] to enter the Alarm Limits menu.
3. Use the programmable keys to set the high and low heart rate limits.
Increases high HR limit
Decreases high HR limit
Increases low HR limit
Decreases low HR limit
Each time you press a key, the corresponding limit changes by 5 bpm. The current HR limits are always shown in the
upper left portion of the display.
4. Press [EXIT] to return to the main menu.
Alarm Type
Default Limit
Heart Rate Low
60
Heart Rate High
120
Pacemaker
Follow the next procedure to activate or deactivate the pacemaker detection function:
1. Press the [SETUP] key from the main menu.
2. Press the [ECG] key and then select the [PACER DETECT] key to toggle between pacer detection enabled or
disabled.
When a pacemaker has been detected, a P will start flashing in the heart symbol.
The message “PACER DETECT DISABLED” will appear if the pacer detection circuit is not active.
WARNING: Rate meters might continue to count the pacemaker rate during occurrences of cardiac arrest or some
arrhythmias. Do not rely on heart rate alarms. Keep pacemaker patients under close surveillance.
HR(BPM
)

Other manuals for Ivy Biomedical Systems 3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ivy Biomedical Systems 3000 Series and is the answer not in the manual?

Ivy Biomedical Systems 3000 Series Specifications

General IconGeneral
BrandIvy Biomedical Systems
Model3000 Series
CategoryMonitor
LanguageEnglish

Related product manuals