JWI-700C 11
4) Wait till “” ”is stable, key in the calibration value
25(1/3 full load) and put the corresponding weights
On, then press key UWS. “25” is flashing.
5) The calibration procedure is completed with a“ ”
flashing on the weight screen. Now, remove all the weights.
6) Press key SMPL to save, and then press key ZERO to return to normal weighing mode.
7. Operation
7-1 Basic Weighing
7-1-1 Weighing
Place item to be weighed on the scale. The Weight
window shown is 2.000kg, Gross Weight.
7-1-2 Tare & Preset Tare
Tare
When weighing a sample that must be held in a container, taring stores the container weight
into memory.
1) Under the weighing mode, place
the container on the scale, weight
window shown is 0.500kg. Press the
Key TARE to complete tare action.
The symbol appears and with
symbol “ ” pointing at “NET”.
2)To clear tare, with an empty pan, Press
down the key TARE or key ZERO.
Preset Tare
1) Under the unload condition , Enter tare
value using the numeric keys together with
key ●, then press key TARE. The weight
window displays the tare value.
2) Put the load on the container, the scale will automatically deduct the value of the container
from the total value.
Note: during the process of Tare or Preset Tare, if the load and container are not removed,
press zero to check the gross weight.