INSTRUCTIONS FOR USE
Arm-type Fully Automatic
Blood Pressure Monitor
DBP-6191

Questions and Answers

  • J
    jennifersalazarAug 10, 2025
    What to do if my JOYTECH Blood Pressure Monitor shows Er5?
    • C
      Chase Payne PhDAug 10, 2025
      If your JOYTECH Blood Pressure Monitor displays 'Er5', it could be due to the pressure exceeding 15 mmHg for 3 minutes, potentially caused by a knotted cuff or blocked vent valve. Check the cuff and valve; if the issue continues, contact the manufacturer. Alternatively, it may be caused by a dead battery, in that case replace the battery or connect the power adapter if available.
  • B
    Bruce MurrayAug 15, 2025
    Why is my JOYTECH DBP-6191 giving abnormal readings?
    • S
      Sarah RossAug 15, 2025
      An abnormal reading on your JOYTECH Blood Pressure Monitor can occur if the armband is either too tight or too loose, or incorrectly positioned. Ensure the armband is correctly rolled. Movement during measurement or interference with the electronic sphygmomanometer can also cause this, so stay still and keep the monitor steady. Speaking, nervousness, or emotional state can affect readings; instead, take deep breaths to calm down. Finally, incorrect posture during measurement can lead to inaccuracies; adjust your posture as described in the “Blood pressure gauge Wearing” section. If you are charging the device or measuring blood pressure improperly, see operation Instructions.
  • R
    Russell OrtizAug 20, 2025
    What does Er3 mean on a JOYTECH DBP-6191 and how to fix it?
    • A
      aguilargeneAug 20, 2025
      If your JOYTECH Blood Pressure Monitor displays 'Er3', it indicates improper compression due to arm or body movement. To resolve this, hold your arm and body still, then measure again.
  • S
    smithkylieAug 25, 2025
    What does Er2 mean on a JOYTECH DBP-6191?
    • S
      Sarah MitchellAug 26, 2025
      If your JOYTECH Blood Pressure Monitor shows 'Er2', it means the cuff is too loose. Please fasten the cuff before measuring to resolve this issue.
  • W
    William SpencerAug 29, 2025
    What does error Er4 mean on JOYTECH DBP-6191?
    • J
      John WheelerAug 29, 2025
      If your JOYTECH Blood Pressure Monitor displays 'Er4', it means that the pressure exceeds 300mmHg. Please fasten the cuff before measuring.
  • J
    jasmine59Sep 3, 2025
    How to fix Er1 error on JOYTECH DBP-6191?
    • P
      pmartinezSep 3, 2025
      If your JOYTECH Blood Pressure Monitor displays 'Er1', it means that the device can’t detect high and low pressure. Please fasten the cuff before measuring to resolve this issue.