|
23ENGLISH
Table 4
Recommended separation distances between portable and mobile RF communications equipment and the device
The device is intended for use in an electromagnetic environment in which radiated therefore disturbances are
controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and the device
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of
transmitter [W]
Separation distance according to frequency of transmitter [m]
80 MHz do 800 MHz 800 MHz do 2,7 GHz
0,01 0,12 0,23
0,1 0,38 0,73
1 1,2 2,3
10 3,8 7,3
100 12
23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
. ADDITIONAL NOTES
Important Instructions Before Use
1.
WARNING: Use of this equipment adjacent to or stacked with other equipment should
be avoided because it could result in improper operation. If such use is necessary,
this equipment and the other equipment should be observed to verify that they are
operating normally.
2.
WARNING: PORTABLE RF communications equipment (including peripherals such
as antenna cables and external antennas) should be used no closer than 30 cm
(12 inches) to any part of Arm-type Fully Automatic Digital Blood Pressure Monitor,
including cables specified by the MANUFACTURER. Otherwise, degradation of the
performance of this equipment could result.