Arm-type Fully Automatic Blood Pressure Monitor DBP-6191
3. The software identifier refer to the software evaluation report , and the file code is
JYRJ210111001.
4. Verify manometer pressure accuracy:
In the power down state, press and hold the “ START/STOP” button, and theninstall
the batteries. Until the LCD screen is full, release the “ START/STOP” button. When the
LCD screen displays the double zero, the bloodpressure meter is in static state. At this
point, 500ml gas capacity, calibrated standard pressure gauge and manual pressure
device can be connected to the sphygmomanometer through the sleeve interface of
the sphygmomanometer, and manual pressure can be applied to the effective display
range of the sphygmomanometer, and then the difference between the reading of
the sphygmomanometer and that of the standard pressure gauge can be compared.
This mode can be used to verify manometer pressure accuracy.
5.
Contraindications: Product is not intended for infants or individuals who cannot
express their intentions.
6. Intended Use: The digital blood pressure monitor are reusable for clinical and home
use and are non-invasive blood pressure measurement systems designed to measure
the systolic and diastolic blood pressure and pulse rate of adolescents and adults
individual by using a non-invasive technique, which is a well-known technique in the
market called the “oscillometric method”. it can measure the systolic blood pressure,
diastolic blood pressure and pulse rated on up-arm,and the device is reusable for
clinical or home use.
7. The patient is the operator:
The PATIENT is an intended OPERATOR.
The PATIENT Do not carry out other maintenance operations except to replace the
battery.
8.
WARNING: Do not modify this equipment without authorization of the manufacturer.
9. ESSENTIAL PERFORMANCE Maintenance advice:
Pressure calibration will be carried out when this product leaves the factory. Patients
can use the method described in the section “Verify Manometer Pressure Accuracy”
to verify the accuracy. If the accuracy deviation is large, please contact the manu-
facturer to recalibration.
10. Mechanical strength and resistance to heat. The resistance to heat will be retained
by device during the EXPECTED SERVICE LIFE of the ME EQUIPMENT
11. Do not place the blood pressure monitor and cuff at will. It will cause asphyxiation
if the child swallows or twine around his neck.