Safety
13
4.3 Safety function
Technical protective devices provide for safe operation. If a safety function is
triggered, the operator is alerted with a message on the display and an acoustic
signal.
Adjustable high temperature cut-off
The high temperature cut-off prevents overheating of the heater.
If the measured temperature rises above the set protective temperature, an
error message is shown on the display. The pump and heater are switched
off. A restart is required.
Overheating protection
The overheating protection prevents overheating of the heater.
The protective mechanism is triggered when the device recognizes a
temperature difference of more than 20 K between the working
temperature sensor and the safety temperature sensor. Am error message
appears on the display. A restart is required.
Low liquid level protection
A level switch recognizes when the bath fluid fill level in the bath tank is too
low. The unit has a warning system to prevent overheating of the heater or dry
running of the pump.
The low liquid level alarm is triggered when the float reaches its lower limit
stop. The device switches off the pump and heater. A continuous signal
tone sounds. An error message appears on the display. A restart is required.