Initial operation
38
Warning! Over-tightening the pump connections may cause the assembly
frame to break.
6. Attach the external system's hoses to the barbed fittings.
Pay attention to the supply and runback position.
7. Using tube clamps, secure the hose to prevent it from slipping.
8. Set the flow direction lever to external circulation.
The external system is connected.
If the external system is disassembled, the pump connections must be
resealed with the sealing plugs so that no bath fluid can splash out during
operation.
7.4 Set high temperature safety function
Before each new temperature application, the temperature must be set for the
tank and internal reservoir high temperature cut-off. Set a value that is at least
25 K below the flash point of the bath fluid being used. The surface temperature
of the bath fluid must not exceed the flash point at any time. An alarm is
triggered when the set value is exceeded.
Slotted screwdriver, size 3
The unit is connected.
1. Switch the unit on.
If no temperature liquid has been filled yet, the low liquid level alarm is
displayed.
Depending on the default setting, the high temperature cut-off alarm is
displayed.