Operation
44
8.6 Record data
8.6.1 Record measurement data
Measurement data for an ongoing temperature control application can also be
recorded onto a USB stick at the same time. The target value temperature,
actual value temperature, and the percentage output are documented every
second in the record. The data is saved as a .txt file and can be subsequently
analyzed.
The unit is switched on.
A USB stick is available.
1. Insert the USB stick into the USB port.
2. Press and hold the [Up Arrow] and [OK] keys simultaneously until the
display shows <LOG I>.
The recording of measurement data is started and indicated on the display
by a flashing dot. Stop recording using the same key combination. The
display will then show <LOG O>.
8.6.2 Reading out blackbox data
The blackbox stores all relevant data from the last 20 minutes. In addition, the
black box logs alarms and warnings.
The blackbox can be read out and the data sent to technical service for analysis.
The unit is switched on.
A USB stick is available.
1. Insert the USB stick into the USB port.
2. Press the [OK] and Service keys simultaneously.
<-BB-> is shown on the display when saving.
The blackbox data is saved onto the USB stick as a .txt file.
8.7 Remote control device
The device can be remote-controlled via the standard USB B interface.
The unit is switched off.
The computer has a terminal program installed.
1. Connect the circulator (USB port type B) to the computer with a standard
USB cable.
2. Download the suitable USB driver from the download area of the website
www.julabo.com.