HE
19
2.5. Cooling water connection
Cooling water pressure (IN/OUT) max. 6 bar
Difference pressure (IN - OUT) 3.5 to 6 bar
Cooling water temperature <20 °C
Recommended quality of cooling water:
pH – value 7,5 to 9,0
Sulfate [SO4 2- ] < 100 ppm
Hydrocarbonate [HCO3- ]/Sulphate [SO4 2-] > 1 ppm
Hardness [Ca2+, Mg2+]/[HCO3-] > 0,5 dH
Alkalinity 60 ppm < [HCO3-] < 300 ppm
Conductivity ȝV/cm
Chloride (CL-) < 50 ppm
Phosphate (PO43-) < 2 ppm
Ammonia (NH3) < 0,5 ppm
Free Chlorine < 0,5 ppm
Ferri Ions (Fe3+) < 0,5 ppm
Mangano Ions (Mn2+) < 0,05 ppm
Carbon dioxide (CO2) < 10 ppm
Hydrosulfide (H2S) < 50 ppm
Content of oxygen < 0,1 ppm
Algae growth impermissible
Suspended solids impermissible
Notice:
Danger of corrosion of heat exchanger due to unsuitable quality of cooling
water.
x Due to its high content of lime hart water is not suitable for cooling and causes
calcination of the heat exchanger.
x Ferrous water or water containing ferrous particles will cause formation of rust
even in heat exchangers made of stainless steel.
x Chlorous water will cause pitting corrosion in heat exchangers made of stainless
steel.
x Due to its corrosive characteristics distilled and deionized water is unsuitable
and will cause corrosion of the bath. .
x Due to its corrosive characteristics sea water is not suitable.
x Due to its microbiological (bacteria) components which settle in the heat
exchanger untreated and unpurified river water and water from cooling towers is
unsuitable.
x Avoid particulate matter in cooling water.
x Avoid putrid water.
Notice: Cooling water circuit
Risk of oil leaking from the cooling circuit (compressor) of the recirculating cooler into
the cooling water in case of a fault in the circuit!
Observe the laws and regulations of the water distribution company valid in the
location where the unit is operated.
www.calcert.com sales@calcert.com1.888.610.7664
0
5
10
15
20
25
30