Menu functions
64
Î
x The parameter flashes, switch by pressing
and .
A >3-point< calibration is carried out.
Î
The value >TMPVAL< is only indicated
In addition the measured value >CALVAL X< is saved
during the following step
Î
x Integer digits flash, set by pressing
(79) +
x Decimal digits flash, set by pressing
(70) +
The first of 3 points is calibrated.
Return to 2. Set working temperature value SETPNT: 120.00 °C
Î
The value is only indicated
Î
x Integer digits flash, set by pressing
(119) +
x Decimal digits flash, set by pressing
(50) +
The second of 3 points is calibrated.
Return to 2. set working temperature value SETPNT: 160.00 °C
Î
The value is only indicated.
Î
x Integer digits flash, set by pressing
(159) +
x Decimal digits flash, set by pressing
(30) +
The 3-point calibration is completed
5. Return to standard operation
Î
x Set >YES< after calibration.
(Standard operation)